A picosecond laser energy stability test does more than confirm that one pulse hit the advertised millijoule figure. A factory acceptance test (FAT) evaluates the measurement equipment, the controlled conditions, repeated-shot output, short-term fluctuation, warm-up drift, spot-size and wavelength behaviour, beam shape and cooling stability — with everything traceable to the exact serial number being shipped. Acceptance criteria must be agreed with the supplier before the test runs, not written after the results are seen.
Quick answer: What does a proper energy stability test cover?
Nine things. A calibrated energy meter with a documented calibration date. Multiple pulse-energy readings — not one hero shot. Repeated-shot variation (min, max, mean, spread). Warm-up drift after a documented interval. Actual spot diameter measured, not assumed. Each installed wavelength tested separately. Beam-profile evidence at more than one spot. Cooling status recorded throughout. And a signed serial-level report linking every result to the specific machine serial number leaving the factory.
How to use this guide
This checklist is for distributors, technical buyers, clinic owners and procurement teams evaluating professional picosecond laser machines before shipment or final payment. Every acceptance tolerance should come from the supplier's documented specification and the signed purchase agreement — not from a generic blog figure.
Picosecond FAT: Test Items, Evidence and Pass/Fail Basis
The framework below identifies each test item, the evidence recorded, why it matters, and the source that defines the pass/fail basis.
PICOSECOND ENERGY-STABILITY FAT — SUMMARY
|
TEST ITEM
|
WHAT TO RECORD
|
WHY IT MATTERS
|
PASS/FAIL BASIS
|
|
Energy meter
|
Model, sensor, calibration date
|
Establishes measurement credibility
|
Approved meter and range
|
|
Pulse energy
|
Multiple readings at defined conditions
|
Verifies actual output
|
Supplier specification
|
|
Repeated-shot variation
|
Min, max, mean, spread
|
Shows stability
|
Agreed tolerance
|
|
Warm-up drift
|
Energy against elapsed time
|
Detects thermal drift
|
Agreed tolerance
|
|
Spot size
|
Actual measured diameter
|
Needed for fluence calculation
|
Specification tolerance
|
|
Wavelength
|
1064/532 nm or installed options
|
Confirms configuration
|
Purchase order
|
|
Beam profile
|
Shape and uniformity report
|
Finds hotspots and asymmetry
|
Agreed acceptance criterion
|
|
Cooling
|
Temperature and any alarms
|
Tests thermal stability
|
No unresolved fault
|
|
Serial report
|
Machine ID linked to results
|
Ties evidence to the shipped unit
|
Required documentation
|
Acceptance limits are set from the supplier's written specification and the purchase agreement — not from a universal figure. Agree every tolerance before the test runs.
What Does Picosecond Laser Energy Stability Actually Mean?
Three properties get confused in supplier conversations. All three must be evaluated separately.
Accuracy = does measured energy match the expected nominal value?
Repeatability = do repeated pulses produce similar energy readings?
Drift = does output change as the machine warms or runs longer?
|
ACCURACY
|
REPEATABILITY
|
|
Closeness to the expected value
|
Closeness of repeated readings to each other
|
|
One reading can look accurate
|
Requires multiple readings
|
|
Does not prove stability
|
Does not automatically prove calibration
|
|
Compare to specification
|
Compare shot-to-shot variation
|
A machine can be repeatable but incorrectly calibrated, or accurate on one pulse but unstable across many pulses. Any FAT that stops after one impressive reading has verified neither.
Factory Acceptance Test vs Site Acceptance Test
A factory acceptance test (FAT) runs before shipment; a site acceptance test (SAT) runs after installation. Both are needed — FAT prevents the wrong configuration shipping, SAT catches transport and installation problems.
|
FACTORY ACCEPTANCE TEST (FAT)
|
SITE ACCEPTANCE TEST (SAT)
|
|
Before shipment
|
After installation
|
|
Factory environment
|
Clinic environment
|
|
Prevents wrong-configuration shipping
|
Finds shipping and setup problems
|
|
Supports final-payment decision
|
Supports clinical handover
|
|
Supplier normally performs, buyer witnesses
|
Buyer or service team verifies
|
Define Pass/Fail Criteria Before the Test Runs
Every acceptance parameter must be written and agreed before the machine is switched on: wavelength, spot size, nominal energy, allowed variation, repetition rate, warm-up requirement, test duration and beam criteria. The most common way FATs fail buyers is when tolerance is set after the readings are in.
-
Do not let the supplier change acceptable variation after seeing results
-
Do not discard a bad reading without a documented reason
-
Do not retest at easier settings to obtain a pass
-
Do not move the measurement location partway through the sequence
Structure the acceptance document with six columns per parameter: parameter, target, tolerance, test method, required evidence, pass/fail.
Calibrated Energy Meter vs Machine Display

The machine display shows the programmed setpoint against stored calibration — it cannot verify itself. An external calibrated energy meter provides the physical measurement the FAT is built on.
|
MACHINE DISPLAY
|
CALIBRATED ENERGY METER
|
|
System setpoint
|
Independent measurement
|
|
Useful during normal operation
|
Useful for acceptance
|
|
Depends on machine calibration
|
Has separate calibration
|
|
Cannot verify itself
|
Provides external evidence
|
Meter suitability
Record meter manufacturer, model, sensor head, wavelength range, energy range, calibration status and serial number. Test equipment must tolerate the pulse energy, peak power, wavelength and repetition rate being measured. Do not improvise neutral-density filters or attenuators without a validated setup.
Set Up a Repeatable Test Condition
A useful FAT requires repeatable geometry, not hand-held measurements that change from pulse to pulse. Record four categories before the first shot.
Environment: room temperature, machine location, airflow around cooling vents, supply voltage where relevant.
Machine state: cold or warmed up, coolant temperature and status, error and warning state, handpiece or delivery system.
Laser setting: wavelength, spot size, energy setting, repetition rate, pulse mode.
Geometry: measurement distance, handpiece orientation, sensor position, spot location on the sensor.
Cold-Start Energy vs Warm-System Energy
A cold-start reading is the initial baseline. A warm-system reading — taken after sustained operation at the tested settings — reveals the thermal behaviour that determines real clinic performance. A test that stops after cold-start alone has skipped the harder half.
|
COLD-START ENERGY
|
WARM-SYSTEM ENERGY
|
|
Initial condition
|
Thermally loaded condition
|
|
Easy to demonstrate
|
More representative of sustained use
|
|
May look stable
|
Can expose drift
|
|
Does not test cooling capacity
|
Tests thermal interaction
|
Short-Term Fluctuation vs Long-Term Drift
Fluctuation is rapid shot-to-shot variation; drift is a gradual movement of the mean over time. The two need different metrics: fluctuation is captured by shot-to-shot spread, drift by a time-series trend.
|
FLUCTUATION
|
DRIFT
|
|
Fast, shot-to-shot changes
|
Slow change over minutes
|
|
May require statistical spread
|
Requires time-series trend
|
|
Can average around the same mean
|
The mean itself moves
|
|
Missed by cold-start test alone
|
Missed by short sequences
|
Number of Shots and Repeated-Shot Variance
There is no universal number of shots. The sequence length is set in the acceptance document based on the supplier specification and the operating configuration. A workable framework: baseline sequence, sustained sequence, warm sequence — each documented separately for each wavelength, spot size and repetition rate.
Do not cherry-pick readings
The raw sequence should be retained in full. If any reading is invalidated, state why in writing, keep the original in the raw data, and identify its replacement.
Base statistics and variation metrics
Record minimum, maximum, mean and range. Add a variation metric — range relative to mean, standard deviation, or coefficient of variation — and use it consistently.
Mean energy = sum of measured pulses ÷ number of pulses recorded.
Coefficient of variation = standard deviation ÷ mean, expressed as a percentage. Compare to the supplier's written stability specification — do not adopt a universal figure from an unrelated product.
|
SUPPLIER SPECIFICATION
|
TEST RESULT
|
|
Contractual or quoted stability
|
Measured sequence
|
|
Defines the acceptance limit
|
Shows whether the limit is met
|
|
Must state test conditions
|
Must use comparable conditions
|
Nominal Energy vs Delivered Energy
Nominal energy is what the interface or specification claims. Delivered energy is what reaches the measurement plane. Ask the supplier where its stated energy is measured — inside the laser cavity, before delivery optics, after the articulated arm or at the treatment exit. The picosecond laser buyer's guide recommends requesting energy at the handpiece and at each spot size.
1064 nm vs 532 nm Energy Testing

Each installed wavelength must be tested separately. Stability at 1064 nm does not prove stability at 532 nm — the wavelengths use different detector settings, and where 532 nm is produced by frequency conversion the output characteristics may differ.
|
1064 NM TEST
|
532 NM TEST
|
|
Separate energy baseline
|
Separate energy baseline
|
|
Separate detector compatibility
|
Separate detector compatibility
|
|
Does not validate 532 nm
|
Does not validate 1064 nm
|
|
Record independently
|
Record independently
|
The C19 laser system product page publishes different maximum energy values for 1064 nm and 532 nm — a working example of why each wavelength needs its own test record and its own agreed tolerance.
Spot Size, Diameter Verification and Beam Uniformity

Energy stability at one spot does not prove stability at other spots. Prioritise smallest practical, typical operating and largest advertised spot. Fluence equals energy divided by spot area — an incorrect spot diameter changes calculated fluence even when the meter reading is correct, which is why diameter must be measured, not assumed.
SPOT-SIZE ENERGY-STABILITY WORKSHEET
|
SPOT SIZE
|
NOMINAL ENERGY
|
MEAN MEASURED
|
VARIATION
|
MEASURED DIAMETER
|
RESULT
|
|
Small
|
Record
|
Record
|
Calculate
|
Measure
|
Pass/fail
|
|
Medium
|
Record
|
Record
|
Calculate
|
Measure
|
Pass/fail
|
|
Large
|
Record
|
Record
|
Calculate
|
Measure
|
Pass/fail
|
Functional worksheet — the buyer records actual values into each cell during the FAT. The C16 and C19 pages advertise spot ranges from 2–10 mm, so "Small / Medium / Large" for those platforms typically means 2 mm, a mid-range spot, and 10 mm.
Spot diameter vs beam uniformity
|
SPOT DIAMETER
|
BEAM PROFILE
|
|
Measures physical footprint
|
Measures energy distribution
|
|
Can be correct despite a hotspot
|
Shows centre-to-edge behaviour
|
|
Needed for fluence
|
Needed for uniformity
|
|
Simple dimensional parameter
|
Optical-quality parameter
|
A visual spot on a target card shows gross centring but cannot quantify uniformity or resolve small hotspots. Where uniformity is part of the acceptance criteria, request a beam-profiler report — a photograph or burn mark is not a substitute for quantitative data.
Repetition Rate and Cooling Stability
Low-rate energy is a baseline. Higher rates test capacitor recharge, thermal loading and cooling capacity. Because the C16 and C19 platforms advertise operation up to 10 Hz, the FAT should document behaviour at the repetition rates in the purchase specification, not only at the easiest setting.
|
LOW HZ TEST
|
HIGH HZ TEST
|
|
Easier thermal condition
|
Higher sustained load
|
|
Useful baseline
|
Tests clinic-speed configuration
|
|
May maximise pulse energy
|
Energy may behave differently
|
|
Does not prove high-rate stability
|
More demanding FAT condition
|
Record cooling during the test
Track coolant temperature, machine temperature where displayed, fan and pump status, flow warnings and thermal warnings throughout the sequence. If output becomes unstable only as the machine heats, a cold-start test will miss it — which is exactly why the FAT includes a warm-system stage.
Factory Demo Video vs Buyer-Grade FAT Evidence

A promotional video shows the machine firing, a meter display and one energy result. Buyer-grade FAT evidence documents compliance.
|
PROMOTIONAL DEMO VIDEO
|
BUYER-GRADE FAT EVIDENCE
|
|
Shows capability
|
Documents compliance
|
|
Often edited
|
Continuous sequence preferred
|
|
May omit conditions
|
Conditions visible throughout
|
|
One impressive reading
|
Repeated dataset
|
What the FAT video and report should show
The video should keep machine model, serial number, test instrument, calibration label, wavelength, spot size, energy setting and repetition rate visible throughout the sequence — with cold-start and warm-system readings recorded in the same take. The written report ties every result back to the specific serial number being shipped. The LEFIS certifications and compliance page describes pre-shipment performance testing and batch traceability, which is the basis for insisting on serial-level FAT evidence.
C16 vs C19: The FAT Fields to Populate
The two LEFIS platforms illustrate what a scoped FAT looks like on a real Q-switched Nd:YAG picosecond-category system. This is not a "which is better" comparison — both should be tested against the same acceptance framework, at their own published specifications.
C16 test fields
The C16 laser system page publishes 1064/532 nm, up to 10 Hz, 2–10 mm spots and closed-loop water-plus-air cooling. FAT questions: what energy does the shipped unit measure at each wavelength? How stable is repeated output? Does stability change across spot sizes? What happens after warm-up? What is the result at higher Hz?
C19 test fields
The C19 laser system page publishes 1064 nm up to 1000 mJ, 532 nm up to 500 mJ, 2–10 mm spots, up to 10 Hz and liquid-plus-forced-air cooling. Populate the same FAT questions per wavelength, using the C19's own published maximum energy values as the supplier specification.
Pass/Fail Report Template
A functional worksheet for the buyer to complete during the FAT. Every Requirement column is populated from the supplier's written specification agreed before testing; Measured result, Pass/Fail and Evidence file columns are completed during the test itself.
Machine and equipment identification
-
Machine: manufacturer, model, serial number, software version, manufacture date
-
Measurement equipment: energy meter, sensor head, serial number, calibration date
-
Test conditions: room temperature, cooling status, warm-up time, wavelength, spot size, repetition rate, nominal energy
FAT RESULTS TABLE
|
TEST
|
REQUIREMENT
|
MEASURED RESULT
|
PASS/FAIL
|
EVIDENCE FILE
|
|
1064 nm energy
|
Supplier spec
|
Record
|
P/F
|
File ref
|
|
532 nm energy
|
Supplier spec
|
Record
|
P/F
|
File ref
|
|
Repeated-shot variation
|
Agreed limit
|
Calculate
|
P/F
|
File ref
|
|
Warm-up drift
|
Agreed limit
|
Calculate
|
P/F
|
File ref
|
|
Spot diameter
|
Supplier spec
|
Measure
|
P/F
|
File ref
|
|
Beam uniformity
|
Agreed criterion
|
Record
|
P/F
|
File ref
|
|
High-Hz stability
|
Agreed criterion
|
Record
|
P/F
|
File ref
|
|
Cooling test
|
No unresolved fault
|
Record
|
P/F
|
File ref
|
Requirement column: populated from the supplier's written specification before the test. Measured result and Pass/Fail: completed during the test. Evidence file: references the video, screenshot or raw data supporting each row.
Retest Triggers and Stop-Acceptance Conditions
Automatically retest when
A sequence should be repeated — with the reason documented — if any of the following occur: meter overload or sensor saturation, machine warning, accidental geometry change, spot-size or wavelength change, cooling fault, handpiece movement, unexplained outlier, or sequence interruption. Retests must document why the first sequence was invalidated rather than silently replacing poor results with better ones.
Hold acceptance pending investigation when
-
Measured output falls outside the agreed specification
-
Repeated instability exceeds the agreed tolerance
-
Warm-up drift persists across a full sequence
-
Wavelength configuration or spot dimensions do not match the purchase order
-
Severe beam asymmetry or hotspot is present
-
Recurring cooling alarms cannot be cleared
-
Output drops unexplained at the rated repetition rate
-
Serial number, meter calibration or factory test record is missing, mismatched or expired
Final Picosecond Laser Energy Stability Test Checklist
Before accepting the machine — or releasing final payment — confirm the FAT has:
-
confirmed the machine serial number and defined pass/fail tolerance from the supplier's written specification
-
verified energy-meter suitability, sensor range and current calibration status
-
recorded environmental conditions, cooling status, wavelength, spot, energy and Hz per sequence
-
recorded cold-start readings, then repeated readings after warm-up
-
calculated mean, variation and warm-up drift for each configuration tested
-
verified actual spot diameter and inspected beam uniformity
-
tested each installed wavelength separately and the applicable higher repetition rates
-
recorded cooling behaviour throughout the sequence
-
saved raw readings and continuous FAT video
-
completed a serial-level pass/fail report against pre-agreed criteria
Request a Factory Acceptance Video and Energy Test Report
One pulse does not establish stability. Nominal energy, measured energy and stability are three separate properties — all three should be evaluated at every installed wavelength, at more than one spot size, at more than one repetition rate, and both cold and warm — with raw data preserved and every result tied to the specific serial number being shipped.
Compare the LEFIS picosecond laser machines and use the laser technical support page to request the continuous FAT video, the serial-level energy test report and pass/fail documentation for the exact configuration on the quotation. Confirm acceptance tolerance and evidence format in writing before the test runs.
FAQs
What is a picosecond laser energy stability test?
A factory acceptance test that verifies measured pulse energy at the handpiece, with calibrated equipment, at defined wavelength, spot size and repetition rate, across repeated readings both cold and warm. The result is compared against the supplier's written specification agreed before testing.
How do you measure picosecond laser pulse energy?
With a calibrated energy meter and a sensor rated for the wavelength, pulse energy, peak power and repetition rate. Record meter and sensor model, serial and calibration date; hold spot size, distance, orientation and machine state constant; take multiple readings, not one hero shot.
How many pulses should be tested during factory acceptance?
There is no universal number. Set sequence length in the acceptance document based on the supplier specification and configuration. A workable framework: baseline sequence, sustained sequence, warm sequence — each recorded separately per wavelength, spot size and repetition rate.
What is the difference between laser fluctuation and drift?
Fluctuation is fast shot-to-shot variation. Drift is a slow change of the mean over time — upward, downward or cyclic. Fluctuation is captured with statistical spread; drift with a time-series trend. A short cold-start test can pass both while missing them entirely.
Why should picosecond laser energy be tested after warm-up?
Because thermal drift, cooling weakness and alignment sensitivity only appear once the machine has been running at operating load. Cold-start verifies the first pulse; warm-system verifies the machine can maintain performance across a treatment session.
Should 532 nm and 1064 nm energy be tested separately?
Yes. Stability at one wavelength does not prove stability at the other. Different detector settings apply, and where 532 nm is produced by frequency conversion the characteristics can differ. Record independent baselines, means, variations and results per wavelength.
Does spot size affect picosecond laser energy testing?
Yes. Fluence equals energy divided by spot area — the same pulse energy at 2 mm produces very different fluence from 10 mm. Most systems do not maintain headline mJ at every listed spot. Test smallest practical, typical operating and largest advertised, and verify actual diameter.
How do you check whether a laser spot is uniform?
With a beam profiler producing quantitative centre-to-edge distribution data at more than one spot size. Burn paper or a target-card mark shows gross centring but cannot quantify uniformity or resolve small hotspots. Agree the uniformity acceptance criterion in writing before the FAT.
What should a picosecond laser factory acceptance report include?
Machine model, serial, software and manufacture date; meter, sensor, serial and calibration date; environmental and cooling conditions; wavelength, spot, Hz and nominal energy per sequence; raw readings, calculated statistics, pass/fail against pre-agreed limits, and file references for video, screenshots and logs.
What happens if a laser fails an energy stability test?
Hold acceptance pending investigation — do not condemn the machine automatically. Common causes: incorrect meter or wavelength setting, geometry change during the sequence, cooling fault, incorrect spot size, or drift after warm-up. Document the failed criterion, retest with reason recorded, and escalate before releasing final payment.
Sources
-
How to Measure Unstable Laser Power · Ophir Photonics
-
Laser Products and Instruments · U.S. Food & Drug Administration
-
IEC 60825-1 — Safety of Laser Products · International Electrotechnical Commission
-
ISO 13485 — Medical Devices Quality Management Systems · International Organization for Standardization
-
Physical Measurement Laboratory — Optical Radiation · U.S. National Institute of Standards and Technology
Looking for something else?
Does Tattoo Removal Leave Scars? Truth, Risks & Recovery
LEARN MORE
Is Tattoo Removal Safe? A Complete Medical & Technology Guide
LEARN MORE
Is Laser Hair Removal Painful? What It Really Feels Like & How to Manage It
LEARN MORE
What Is IPL Hair Removal? The Full Guide to Technology. Safety. and Results
LEARN MORELooking for something else?
Can You Do Laser Hair Removal While Pregnant? Safety, Risks & Medical Guidance
LEARN MORE
How to Prepare for Laser Hair Removal: The Ultimate Step-by-Step Guide
LEARN MORE
How to Remove Facial Hair for Women: A Complete Guide to Safe. Smooth Skin
LEARN MOREYou may also like
Further reading
How to Choose CO2 Laser Tube Power and Lifespan for Your Clinic Workload
CO2 laser tube power and lifespan are the two specifications aesthetic-clinic buyers most often misread. Headline wattage is not proof the source will meet a busy fractional and surgical mix; a rat...
How to Test Picosecond Laser Energy Stability Before Accepting a Machine
A picosecond laser energy stability test does more than confirm that one pulse hit the advertised millijoule figure. A factory acceptance test (FAT) evaluates the measurement equipment, the contro...
Picosecond Laser Machine Maintenance: Optics, Water, Handpiece and Energy Output Testing
Picosecond laser machine maintenance is about output stability, not just a clean casing. Four things determine whether the platform performs the same today as it did on day one: optical cleanliness...
Picosecond Laser Machine Supplier Checklist: Pulse Energy, Spot Size, Wavelengths and Service
A picosecond laser machine supplier should prove — with test data, not brochure copy — that the platform delivers the pulse duration, energy, spot-size range, wavelength coverage and beam quality i...